Product Description
58L-HCL-25 | X02NI99CP58H821 | 25 PPM Hydrogen Chloride | Balance Nitrogen | 58 Liter | connection C-10 | Certificate NIST, | Made in the USA |
The Calgaz X02NI99CP58H821 (Calgaz Part # X02NI99CP58H821 / Cross-Reference # 58L-HCL-25) is an industry-grade, NIST-traceable 25 PPM Hydrogen Chloride (HCl) with a Balance of Nitrogen (N2) specialty calibration gas mixture. Packaged in a portable, high-integrity 58-liter aluminum cylinder (Calgaz 6D), this reactive standard is designed for precision span calibration, routine bump testing, and field response verification of industrial toxic gas detectors.
Mixture & Compliance Highlights
- Target Analyte: 25 PPM Hydrogen Chloride (HCl)
- Balance Gas: High-Purity Nitrogen (N2)
- Traceability: Certified NIST-traceable standard blended under strict analytical tolerances
- Instant COA Access: Official Certificate of Analysis (COA) is accessible directly on demand via the cylinder's scannable on-bottle QR code
- Direct Cross-Reference: Matches OEM / industry part # 58L-HCL-25
- Cylinder Model: Calgaz 58 Liter Seamless Aluminum
- Origin: Precision blended and certified in the USA
Technical Specifications
| Manufacturer | Calgaz (an Air Liquide company) |
|---|---|
| Part Numbers | X02NI99CP58H821 | 58L-HCL-25 |
| Gas Mixture | 25 PPM Hydrogen Chloride (HCl) / Balance Nitrogen (N2) |
| Cylinder Capacity | 58 Liters (Calgaz 6D) |
| Cylinder Construction | Seamless Aluminum (Passivated interior for reactive gas stability) |
| Outlet Connection | C-10 (5/8"-18 UNF Thread) |
| Working Pressure | Approx. 500 psig (34.5 bar) @ 70°F |
| Analytical Tolerance | ±10% NIST Traceable |
| Certificate of Analysis | Included via Scannable On-Cylinder QR Code |
| Warranty / Shelf Life | 12 Months (Reactive Gas Standard) |
| Country of Origin | Made in the USA |
Hardware & Handling Requirements
Hydrogen Chloride (HCl) is an aggressive, moisture-reactive acid gas that requires dedicated, non-reactive sampling hardware:
- Regulator Compatibility: Use with a C-10 compatible regulator constructed from Stainless Steel or passivated Nickel-Plated Brass (e.g., Calgaz 70-SS / 714 / 700 series fixed flow or Calgaz DFR 2000 series reactive demand flow regulators). Standard brass regulators are not recommended for HCl service.
- Sample Line Tubing: Strictly utilize PTFE or FEP sample tubing. Never use Tygon, rubber, or standard polyurethane tubing, which will rapidly strip reactive HCl molecules from the stream, resulting in false low readings.
- Instrument Applicability: Fully compatible with electrochemical HCl sensors across leading gas detection platforms including Honeywell Analytics, BW Technologies, RAE Systems, MSA Safety, Dräger, and Industrial Scientific.
Product Overview & Cross-Reference: Calgaz part number 58L-HCL-25 (cross-referenced to Gasco part number X02NI99CP58H821 is a certified NIST-traceable calibration gas standard containing 25 PPM Hydrogen Chloride (HCl) in a balance of Nitrogen (N2). Blended in a 58-liter aluminum cylinder (Calgaz 6D) rated at 500 psig with a standard C-10 valve connection (5/8"-18 UNF), this reactive calibration gas is manufactured in the United States and includes an on-demand Certificate of Analysis (COA) accessible directly via on-cylinder QR code. Engineered for bump testing and span calibration of fixed and portable toxic gas detection systems from BW Technologies, RAE Systems, MSA, Dräger, and Industrial Scientific.